EasyManua.ls Logo

GW Instek GPT-9500 Series - [:Source]:Safety:step<N>:Osc:channel[:High]; [:Source]:Safety:step<N>:Osc:channel:low

GW Instek GPT-9500 Series
283 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
259
[:SOURce]:SAFEty:STEP<n>:OSC:CHANnel[:HIGH]
Sets the output channel status for OSC of scanning test high
voltage.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))
Example: SAFE:STEP4:OSC:CHAN (@(1))
Sets the output channel for OSC of step4 to channel 1 HIGH
output
Example: SAFE:STEP4:OSC:CHAN (@(0))
Sets the output channel for OSC of step4 off for all HIGH
output.
[:SOURce]:SAFEty:STEP<n>:OSC:CHANnel[:HIGH]?
Returns the output channel status for OSC of scanning test
high voltage.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: (@(CH))
Example: SAFE:STEP4:OSC:CHAN?
> (@(1))
The output channel for OSC of step4 is set to channel 1 HIGH
output.
[:SOURce]:SAFEty:STEP<n>:OSC:CHANnel:LOW
Sets the output channel status for OSC of scanning test
return(low).