GPT-9500 Series User Manual
248
The output channel for DCW of step2 is set to channel 2 and 4
return.
[:SOURce]:SAFEty:STEP<n>:DC:REF
Sets the test offset value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:REF 0.00001
Sets the test offset value for DCW of step2 to 0.01mA.
[:SOURce]:SAFEty:STEP<n>:DC:REF?
Returns the test offset value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP2:DC:REF?
> +1.000000E-05
The test offset value for DCW of step2 is 0.01mA.
[:SOURce]:SAFEty:STEP<n>:DC:GROUndmode
Sets the ground mode for DCW in selected step to on or off.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: 0 | 1 | OFF | ON
Example: SAFE:STEP2:DC:GROU 0
Sets the ground mode for DCW of step2 to off.