REMOTE CONTROL
265
[:SOURce]:SAFEty:PRESet:WRANge[:AUTO]
Sets the auto range function for DCW on or off.
Parameter: 0 | 1 | OFF | ON
Example: SAFE:PRES:WRAN 1
Sets the auto range function for DCW on.
[:SOURce]:SAFEty:PRESet:WRANge[:AUTO]?
Returns the status of auto range function for DCW.
Return parameter: 0 | 1
Example: SAFE:PRES:WRAN?
> 1
The status of auto range function for DCW is on.
[:SOURce]:SAFEty:PRESet:GFI[:SWITch]
Sets the GFCI switch on or off.
Parameter: 0 | 1 | OFF | ON
Example: SAFE:PRES:GFI 0
Sets the GFCI switch off.
[:SOURce]:SAFEty:PRESet:GFI[:SWITch]?
Returns the GFCI switch status.
Return parameter: 0 | 1
Example: SAFE:PRES:GFI?
> 0
The GFCI switch status is off.