> +5.000000E-01
The ramp time value for IR of step3 set is 0.5s.
[:SOURce]:SAFEty:STEP<n>:IR:TIME[:TEST]
Sets the test time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP3:IR:TIME 5
Sets the test time value for IR of step3 to 5s.
[:SOURce]:SAFEty:STEP<n>:IR:TIME[:TEST]?
Returns the test time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP3:IR:TIME?
> +5.000000E+00
The test time value for IR of step3 set is 5s.
[:SOURce]:SAFEty:STEP<n>:IR:TIME:FALL
Sets the fall time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP3:IR:TIME:FALL 0
Sets the fall time value for IR of step3 to off.
[:SOURce]:SAFEty:STEP<n>:IR:TIME:FALL?
Returns the fall time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)