OPERATION
59
Setting the Wait Time
Background
The Wait Time refers to the pending time
before judgment appears. Generally, test time
begins after ramp up time; however, a wait
time can be intervened between ramp up time
and test and begins depending on relevant
settings. See page 159 for details.
The Wait Time is applicable for ACW, DCW
and IR tests.
Steps
1. Press the MANU/AUTO soft-key
to select MANU option.
2. Press the EDIT soft-key followed by
clicking the PAGE key for twice.
x 2
3. Press the WAIT TIME soft-key
followed by using the scroll wheel
to set the WAIT TIME value.
ACW
DCW
IR
OFF, 000.1s~999.9s
OFF, 000.1s~999.9s
OFF, 000.1s~999.9s