REMOTE CONTROL
255
[:SOURce]:SAFEty:STEP<n>:IR:CHANnel[:HIGH]
Sets the output channel status for IR of scanning test high voltage.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))
Example: SAFE:STEP3:IR:CHAN (@(1,3))
Sets the output channel for IR of step3 to channel 1 and 3
HIGH output
Example: SAFE:STEP3:IR:CHAN (@(0))
Sets the output channel for IR of step3 to off for all HIGH
output.
[:SOURce]:SAFEty:STEP<n>:IR:CHANnel[:HIGH]?
Returns the output channel status for IR of scanning test high
voltage.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: (@(CH))
Example: SAFE:STEP3:IR:CHAN?
> (@(1,3))
The output channel for IR of step3 is set to channel 1 and 3
HIGH output.
[:SOURce]:SAFEty:STEP<n>:IR:CHANnel:LOW
Sets the output channel status for IR of scanning test return (low).
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))
Example: SAFE:STEP3:IR:CHAN:LOW (@(2,4))
Sets the output channel for IR of step3 to channel 2 and 4 return.