REMOTE CONTROL
253
Return Parameter: <NR3>
Example: SAFE:STEP3:IR:TIME:FALL?
> +0.000000E+00
The fall time value for IR of step3 set is off.
[:SOURce]:SAFEty:STEP<n>:IR:RANGe[:UPPer]
It is in accordance with users’ input current value to select the
range which is upper than the current that can be measured.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP3:IR:RANG 0.0005
The IR measured current value of step3 is set to 500uA. Thus,
meanwhile the selected IR range upper than the current can be
measured is 5mA.
[:SOURce]:SAFEty:STEP<n>:IR:RANGe[:UPPer]?
Returns the range for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP3:IR:RANG?
> +5.000000E-03
The setting range for IR of step3 is 5mA.
[:SOURce]:SAFEty:STEP<n>:IR:RANGe:LOWer
It is in accordance with users’ input current value to select the
range which is lower than the current that can be measured.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>