REMOTE CONTROL
247
[:SOURce]:SAFEty:STEP<n>:DC:CHANnel[:HIGH]?
Returns the output channel status for DCW of scanning test
high voltage.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: (@(CH))
Example: SAFE:STEP2:DC:CHAN?
> (@(1,3))
The output channel for DCW of step2 is set to channel 1 and 3
HIGH output.
[:SOURce]:SAFEty:STEP<n>:DC:CHANnel:LOW
Sets the output channel status for DCW of scanning test
return(low).
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))
Example: SAFE:STEP2:DC:CHAN:LOW (@(2,4))
Sets the output channel for DCW of step2 to channel 2 and 4 return.
Example: SAFE:STEP2:DC:CHAN:LOW (@(0))
Sets the output channel for DCW of step2 to off for all return.
[:SOURce]:SAFEty:STEP<n>:DC:CHANnel:LOW?
Returns the output channel status for DCW of scanning test
return(low).
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: (@(CH))
Example: SAFE:STEP2:DC:CHAN:LOW?
> (@(2,4))