OPERATION
61
Setting the Fall Time
Background
The FALL time is the time taken for the DUT to
discharge the test voltage level. The FALL time
can be set from OFF to 999.9 seconds. The FALL
time is applicable for ACW, DCW and IR tests.
Output Voltage
Timing Chart
(Resistive load)
Test V
time
TEST TIME
RAMP UP
FALL
Steps
1. Press the MANU/AUTO soft-key
to select MANU option.
2. Press the EDIT soft-key followed by
clicking the PAGE key for twice.
x 2
3. Press the FALL TIME soft-key
followed by using the scroll wheel
to set the FALL TIME value.
ACW
DCW
IR
OFF, 000.1s~999.9s
OFF, 000.1s~999.9s
OFF, 000.1s~999.9s