[:SOURce]:SAFEty:STEP<n>:IR:TIME:DWELl
Sets the wait time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP3:IR:TIME:DWEL 0.9
Sets the wait time value for IR of step3 to 0.9s.
[:SOURce]:SAFEty:STEP<n>:IR:TIME:DWELl?
Returns the wait time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP3:IR:TIME:DWEL?
> +9.000000E-01
The wait time value for IR of step3 set is 0.9s.
[:SOURce]:SAFEty:STEP<n>:IR:TIME:RAMP
Sets the ramp time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP3:IR:TIME:RAMP 0.5
Sets the ramp time value for IR of step3 to 0.5s.
[:SOURce]:SAFEty:STEP<n>:IR:TIME:RAMP?
Returns the ramp time value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP3:IR:TIME:RAMP?