EasyManuals Logo

GW Instek GPT-9500 Series User Manual

GW Instek GPT-9500 Series
283 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #250 background imageLoading...
Page #250 background image
GPT-9500 Series User Manual
250
Example: SAFE:STEP3:IR:LIM:HIGH 50000000000
Sets the high limit value for IR of step3 to 50GΩ.
[:SOURce]:SAFEty:STEP<n>:IR:LIMit:HIGH?
Returns the high limit value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP3:IR:LIM:HIGH?
> +5.000000E+10
The high limit value for IR of step3 is 50GΩ.
[:SOURce]:SAFEty:STEP<n>:IR:LIMit[:LOW]
Sets the low limit value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP3:IR:LIM 100000
Sets the low limit value for IR of step3 to 0.1MΩ.
[:SOURce]:SAFEty:STEP<n>:IR:LIMit[:LOW]?
Returns the low limit value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP3:IR:LIM?
> +1.000000E+05
The low limit value for IR of step3 is 0.1MΩ.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9500 Series and is the answer not in the manual?

GW Instek GPT-9500 Series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9500 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals