EasyManua.ls Logo

GW Instek GPT-9500 Series - [:Source]:Safety:step<N>:DC:limit:low; [:Source]:Safety:step<N>:DC:limit:arc[:Level]

GW Instek GPT-9500 Series
283 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
243
Return parameter: <NR3>
Example: SAFE:STEP2:DC:LIM?
> +9.000000E-03
The leakage current high limit for DCW of step2 is 9mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:LOW
Sets the leakage current low limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:LIM:LOW 0.0001
Sets the leakage current low limit for DCW of step2 to 0.1mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:LOW?
Returns the leakage current low limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP2:DC:LIM:LOW?
> +1.000000E-04
The leakage current low limit for DCW of step2 is 0.1mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:ARC[:LEVel]
Sets the ARC value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:LIM:ARC 0.006
Sets the ARC value for DCW of step2 to 6mA.