GW Instek GPT-9500 Series - [:Source]:Safety:step<N>:Ir:channel:low; [:Source]:Safety:step<N>:Ir:ref

GW Instek GPT-9500 Series
283 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPT-9500 Series User Manual
256
Example: SAFE:STEP3:IR:CHAN:LOW (@(0))
Sets the output channel for IR of step3 to off for all return.
[:SOURce]:SAFEty:STEP<n>:IR:CHANnel:LOW?
Returns the output channel status for IR of scanning test return
(low).
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: (@(CH))
Example: SAFE:STEP3:IR:CHAN:LOW?
> (@(2,4))
The output channel for IR of step3 is set to channel 2 and 4 return.
[:SOURce]:SAFEty:STEP<n>:IR:REF
Sets the test offset value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP3:IR:REF 100000
Sets the test offset value for IR of step3 to 0.1MΩ.
[:SOURce]:SAFEty:STEP<n>:IR:REF?
Returns the test offset value for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP3:IR:REF?
> +1.000000E+05
The test offset value for IR of step3 is 0.1MΩ.