[:SOURce]:SAFEty:STEP<n>:PAuse:UTSIgnal?
Returns the status of UNDER TEST SIGNAL in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: 0 | 1
Example: SAFE:STEP5:PA:UTSI?
> 1
The status of UNDER TEST SIGNAL of step5 is on.
[:SOURce]:SAFEty:STEP<n>:PAuse:TIME[:TEST]
Sets the test needed time for PA in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP5:PA:TIME 5
Sets the test needed time for PA of step5 to 5sec.
[:SOURce]:SAFEty:STEP<n>:PAuse:TIME[:TEST]?
Returns the test needed time for PA in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return paramter: <NR3>
Example: SAFE:STEP5:PA:TIME?
> +5.000000E+00
The test needed time for PA of step5 is 5sec.