REMOTE CONTROL
257
[:SOURce]:SAFEty:STEP<n>:IR:GROUndmode
Sets the ground mode for IR in selected step on or off.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: 0 | 1 | OFF | ON
Example: SAFE:STEP3:IR:GROU 0
Sets the ground mode for IR of step3 off.
[:SOURce]:SAFEty:STEP<n>:IR:GROUndmode?
Returns the ground mode status for IR in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: 0 | 1
Example: SAFE:STEP3:IR:GROU?
> 0
The ground mode for IR of step3 is off.
[:SOURce]:SAFEty:STEP<n>:OSC:LIMit:OPEN
Sets the percentage of OSC in selected step judged by open
circuit as detecting short/open circuit.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP4:OSC:LIM:OPEN 0.4
Sets the percentage of OSC of step4 to 40% by open circuit as
detecting short/open circuit.