GW Instek GPT-9500 Series - [:Source]:Safety:step<N>:DC[:Level]; [:Source]:Safety:step<N>:DC:limit[:High]

GW Instek GPT-9500 Series
283 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPT-9500 Series User Manual
242
> 0
The ground mode for ACW of step1 is off.
[:SOURce]:SAFEty:STEP<n>:DC[:LEVel]
Sets the test voltage value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NR1> (50 ~ 6000)
Example: SAFE:STEP2:DC:LEV 5000
Sets the test voltage value for DCW of step2 to 5000V.
[:SOURce]:SAFEty:STEP<n>:DC[:LEVel]?
Returns the test voltage value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP2:DC:LEV?
> +5.000000E+03
The test voltage value for DCW of step2 is 5000V.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit[:HIGH]
Sets the leakage current high limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:LIM 0.009
Sets the leakage current high limit for DCW of step2 to 9mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit[:HIGH]?
Returns the leakage current high limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)