COMMISSIONING
Part No.: 4417340 Revision 1 Service Manual 65
954 SmartServo FlexLine
The TII-SRV firmware supports the following configurable
compensations:
• Wire weight compensation
• Mechanical hysteresis elimination
• Magnet temperature compensation
• Tank shell temperature compensation for high tanks
• Compensation for wire expansion over temperature
• Wire elongation compensation for free wire
• Wire elongation compensation for wire on drum
• Density compensation for level measurement
• Hydrostatic deformation compensation
• Unbalance compensation
• Roof Immersion Compensation
• Drum temperature compensation
• Displacer volume compensation for pressure for density
measurement
• Displacer volume compensation for temperature for density
measurement
• Tilt compensation for density measurement
For correct functioning of the TII-SRV module in the instrument, the
following entities can be set by using either Engauge or some by
SmartView, summed up per Engauge tab.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell FlexLine SmartServo 954 and is the answer not in the manual?

Honeywell FlexLine SmartServo 954 Specifications

General IconGeneral
BrandHoneywell
ModelFlexLine SmartServo 954
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals