COMMISSIONING 
 
 
84  Service Manual    Part No.: 4417340 Revision 1 
  954 SmartServo FlexLine
 
Name  Value Range  Default Value  Explanation 
[Water density dip]      This command first searches 
for the water level and after 
that performs a density 
measurement upwards. And 
then returns to the default 
operation mode. 
[Density water dip]      This command first performs 
a density measurement 
downwards and after that 
searches for the water level. 
And then returns to the 
default operation mode. 
[Density 
compensations] 
<enable> 
<disable> 
<enable> 
<enable> 
<disable> 
<enable> 
 
Enable or disable the 
following compensations for 
the density: 
-Tilt 
Calibrates the density 
measurement for tilt of the 
gauge 
-Unbalance 
Compensates density for the 
unbalance of the drum 
-Unused 
-Volume 
Compensates density for 
hydrostatic deformation of 
the displacer