EasyManuals Logo

Honeywell touchpoint Pro User Manual

Honeywell touchpoint Pro
241 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
MAN0923_Issue 3_04/15 Touchpoint Pro
Operating Instructions
89
6. Commissioning
5. Select CCB Management and click on Conguration Backup
6. Thesystemwilldisplayamessage“Thecongurationbackupprocessmighttakeafewminutes.
The Local UI is not accessible during this time. Do you want to proceed?”
7. Oncetheactionisconrmed,thecongurationbackupprocessstartsandaprogressmessageis
displayed. The backup may take several minutes depending on the system size
8. TheCongurationbackupleisstoredontheSDCardin“\\HardDisk\\TouchpointPro
Conguration\\TouchpointProCong.bin”
Note: Touchpoint Pro always uses the same lename for conguration backups, therefore it is recommended
that a dated copy of the backup le is kept in another location
6.16 Calibrate mV input channels
WARNING
Channel calibration is compulsory after changing the bead current of an AIM mV channel, and
the channel stays in Inhibit until this is done.
Caution: Gas readings may be generated during this procedure. Disable or inhibit any operations or
actions that may result from gas readings before starting this test
Note: Engineer level access is required
Note: If you wish to enable a calibration reminder for the channel, please edit this setting before starting cali-
bration (see Commissioning, Editing a Congured Channel and Congurable Parameter Reference Guide,
Calibration Reminder Settings)
For catalytic bead type detectors, the Touchpoint Pro Analogue Input Module forms part of the measuring
circuit. Therefore all commissioned mV channels remain in Inhibit state until they are calibrated.
Touchpoint Pro monitors the ageing of catalytic bead detectors and can provide a warning when a sensor
should be replaced. Please select First Spanwhencalibratinganeworreplacementsensorfortherst
time, and Span for all calibrations thereafter. When the sensitivity of the sensor has dropped below 50% of
the original value, a “Sensor Lifetime Expired” warning will be generated during calibration.
Caution: If First Span is not used, the system will compare the sensitivity of the sensor to the default
value of 50 mV. This may result in the “Sensor Lifetime Expired” warning being generated prematurely
or not at all.
To calibrate a mV input channel, please follow the procedure below in combination with the relevant
instructions in the sensor manual.

Table of Contents

Other manuals for Honeywell touchpoint Pro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell touchpoint Pro and is the answer not in the manual?

Honeywell touchpoint Pro Specifications

General IconGeneral
BrandHoneywell
Modeltouchpoint Pro
CategoryControl Systems
LanguageEnglish

Related product manuals