5-4
Hardkey and Softkey Reference
AM
AM Depth Couple Off On
This softkey toggles the AM depth coupling on and off. AM depth coupling links
the AM depth values of
AM Path 1 and AM Path 2. When the values are coupled,
any change you make to one AM depth value is applied to both AM depth values.
Softkey Location: Press
AM, AM Depth Couple Off On
Status after Normal Preset: Off
SCPI Commands:
[:SOURce]:AM[1]|2[:DEPTh]:TRACk ON|OFF|1|0
[:SOURce]:AM[1]|2[:DEPTh]:TRACk?
AM Dual-Sine Ampl Ratio
Press this softkey to set the ratio of the depth of AM tone 2 to the depth of AM
tone 1 as a percent. The range of values allowed is 0 to 100%. The minimum
increment allowed is 0.1%. The sum of the two AM depths results in the total AM
depth (set by the
AM Depth softkey in the initial AM Menu.
For example, if the total AM depth is 30% and you set
AM Dual-Sine Ampl Ratio to
50%, then the depth of AM tone 1 will be 20% and the depth of AM tone 2 will be
10% (50% of 20%) so that the depth of the two tones adds up to 30% (20% plus
10%).
Softkey Location: Press
AM, AM Waveform, Dual-Sine, AM Dual-Sine Ampl Ratio
SCPI Commands:
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent
<val><unit>
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent?