5-6
Hardkey and Softkey Reference
AM
Softkey Location: Press AM, AM Path 1 2
Status after Normal Preset: AM Path 1
SCPI Commands:
[:SOURce]:AM:WIDeband:STATe ON|OFF|1|0
[:SOURce]:AM:WIDeband:STATe?
AM Rate
Use this softkey to change the internal modulation frequency for the AM Path 1
and
AM Path 2 configurations. To change the value, press the softkey once. The
current value for AM rate is displayed in the active entry area. To enter a new
value, rotate the front panel knob until the desired value is displayed, use the up
and down arrow keys, or enter the value using the numeric keypad and press the
Hz or kHz terminator softkey. The range of values allowed is 0.1 Hz to 10 kHz.
(0.1 Hz to 50 kHz is the range allowed if sinewave is selected as the internal
waveform.) The minimum increment allowed is 0.1 Hz. Notice that the new value
of AM rate applies only to whichever AM configuration (
AM Path 1 or AM Path 2)
you have currently selected.
Softkey Location: Press
AM, AM Rate
Status after Normal Preset: 400 Hz
SCPI Commands:
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency <val><unit>
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency?