5-153
Hardkey and Softkey Reference
Sweep/List
SCPI Commands:
[:SOURce]:LIST:TRIGger:SOURce EXTernal
[:SOURce]:LIST:TRIGger:SOURce?
:TRIGger[:SEQuence]:SOURce EXTernal
:TRIGger[:SEQuence]:SOURce?
:TRIGger[:SEQuence]:SLOPe?
Freq
Press this softkey to set the signal generator to sweep frequency data only. The
amplitude is set at a constant level determined by the
Amplitude front panel key.
You can define a sweep containing both amplitude and frequency information and
still choose to sweep frequency only.
Softkey Location: Press
Sweep/List, Sweep, Freq
Status after Normal Preset: Sweep is set to Off
SCPI Commands:
[:SOURce]:FREQuency:MODE LIST
[:SOURce]:FREQuency:MODE?
Freq&Ampl
Press this softkey to set the signal generator to sweep both frequency and
amplitude data. Both the frequency and the amplitude lists must have the same
number of points or they must have only one point. If a list has only one point,
then that point is used for each of the other list’s points.
Softkey Location: Press
Sweep/List, Sweep, Freq&Ampl
Status after Normal Preset: Sweep is set to Off
SCPI Commands:
[:SOURce]:POWer:MODE LIST
[:SOURce]:POWer:MODE?
[:SOURce]:FREQuency:MODE LIST
[:SOURce]:FREQuency:MODE?