5-10
Hardkey and Softkey Reference
AM
AM Sweep Trigger
Pressing this softkey reveals a menu of choices for triggering swept-sine
amplitude modulation. You can choose triggering that occurs immediately,
triggering that is supplied by the HP-IB, triggering on either the positive or
negative edge of a signal supplied to the TRIGGER IN connector, or triggering by
the front panel
Trigger key.
Softkey Location: Press
AM, AM Waveform, Swept-Sine, AM Sweep Trigger
Status after Normal Preset: Immediate
SCPI Commands:
:TRIGger[:SEQuence]:SOURce?
AM Tone 1 Rate
Press this softkey to set the internal modulation frequency for the AM tone 1.
Enter the desired rate using the up and down arrow keys, the front panel knob, or
enter a value using the numeric keypad and press the
kHz or Hz terminator softkey.
The range of values allowed is 0.1 Hz through 50 kHz. The minimum increment
allowed is 0.1 Hz.
This softkey is coupled to the
AM Rate softkey in the initial AM Menu and the
AM Start Rate softkey in the AM Swept-Sine Menu. Any value set for
AM Tone 1 Rate is reflected in the AM Rate and AM Start Rate softkeys. Conversely,
if you change the value for either the
AM Rate softkey or the AM Start Rate softkey,
the value for
AM Tone 1 Rate is changed to match.
Softkey Location: Press
AM, AM Waveform, Dual-Sine, AM Tone 1 Rate
Status after Normal Preset: 400.0 Hz
SCPI Commands:
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency <val><unit>
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency?