5-130
Hardkey and Softkey Reference
Pulse
Pulse Rate
Use this softkey to change the pulse rate for internal square pulse modulation. To
change the rate, press the softkey once. The current value for pulse rate is
displayed in the active entry area. To enter a new value, rotate the front panel knob
until the desired value is displayed, use the up and down arrow keys, or enter the
value using the numeric keypad and press the
Hz or kHz terminator softkey. The
range of values allowed is 0.1 Hz to 50 kHz. Notice that the new value of pulse
rate applies only to internal square pulse modulation; when external pulse
modulation or internal pulse modulation is selected, this softkey is disabled.
Softkey Location: Press
Pulse, Pulse Rate
Status after Normal Preset: 400 Hz
SCPI Commands:
[:SOURce]:PULM:INTernal[1]:FREQuency <val><unit>
[:SOURce]:PULM:INTernal[1]:FREQuency?
Pulse Width
Use this softkey to change the pulse width for internal pulse modulation. To
change the pulse width, press the softkey once. The current value for pulse width
is displayed in the active entry area. To enter a new value, rotate the front panel
knob until the desired value is displayed, use the up and down arrow keys, or enter
the value using the numeric keypad and press the
µsec or msec terminator softkey.
The range of values allowed is 8
µs to 30 s. Notice that the new value of pulse
width applies only to internal pulse modulation; when external pulse modulation
or internal square pulse modulation is selected, this softkey is disabled.
Softkey Location: Press
Pulse, Pulse Width
Status after Normal Preset: 40 µs
SCPI Commands:
[:SOURce]:PULM:INTernal[1]:PWIDth <val><unit>
[:SOURce]:PULM:INTernal[1]:PWIDth?