5-190
Hardkey and Softkey Reference
Utility
State
This softkey lets you specify that only instrument state files are to be listed on the
display when
Memory Catalog is selected.
Catalog type is a persistent state; it is not changed by an instrument preset or by a
power cycle.
Softkey Location: Press
Utility, Memory Catalog, Catalog Type, State
SCPI Commands:
:MEMory:CATalog:STATe?
Step/Knob Ratio
This softkey sets the ratio between the increment value of the step keys and the
front panel knob. For example, if you set the increment value for frequency to
1 MHz and set the step/knob ratio to 10/1, when frequency is the active function
the up and down arrow keys will change the frequency in 1 MHz steps and the
front panel knob will change the frequency in 1/10th of 1 MHz steps. This ratio
only applies when you have a function active that has an associated increment
value.
To change the step/knob ratio, press
Utility, Instrument Adjustments,
Step/Knob Ratio. The number displayed in the active entry area represents the
arrow keys step size in the ratio. To enter a new value, rotate the front panel knob
until the desired value is displayed, use the up and down arrow keys, or enter the
value using the numeric keypad and press the
Enter terminator softkey.
The step/knob ratio is a persistent state; it is not affected by an instrument preset
or by a power cycle.
Softkey Location: Press
Utility, Instrument Adjustments, Step/Knob Ratio
Store Ref Osc Setting
This softkey is provided for adjusting the internal reference oscillator. This
adjustment is for use by service personnel only. Refer to the calibration guide for
more information.
Softkey Location: Press
Utility, Instrument Adjustments,
Reference Oscillator Adjustment, StoreRefOscSetting