5-80
Hardkey and Softkey Reference
LF Out
LF Out Width
Use this softkey to change the pulse width for the pulsed low frequency
waveform. To change the width, press the softkey once. The current value for
pulse width is displayed in the active entry area. To enter a new value, rotate the
front panel knob until the desired value is displayed, use the up and down arrow
keys, or enter the value using the numeric keypad and press the
µsec or msec
terminator softkey. The range of values allowed is from 8
µs to a maximum of the
value for the pulse period.
This softkey is only active when you have selected the internal source to operate
as a function generator and when the LF Out waveform is specified as pulse.
Softkey Location: Press
LF Out, LF Out Source, Function Generator,
LF Out Waveform, More 1 of 2, Pulse, LF Out Width
Status after Normal Preset: 40 µs
SCPI Commands:
[:SOURce]:LFOutput:FUNCtion:PWIDth <val><unit>
[:SOURce]:LFOutput:FUNCtion:PWIDth?
Noise
This softkey lets you specify noise as the waveform for the LF output signal.
This softkey is only active when you have selected the internal source to operate
as a function generator.
Softkey Location: Press
LF Out, LF Out Source, Function Generator,
LF Out Waveform, More 1 of 2, Noise
Status after Normal Preset: LF Out Waveform is set to Sine
SCPI Commands:
[:SOURce]:LFOutput:FUNCtion:SHAPe NOISe
[:SOURce]:LFOutput:FUNCtion:SHAPe?

Table of Contents

Other manuals for HP ESG Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP ESG Series and is the answer not in the manual?

HP ESG Series Specifications

General IconGeneral
BrandHP
ModelESG Series
CategoryPortable Generator
LanguageEnglish

Related product manuals