5-152
Hardkey and Softkey Reference
Sweep/List
Status after Normal Preset: List
Softkey Location: Press
Sweep/List, Configure List Sweep, More (1 of 2),
Dwell Type List Step
SCPI Commands:
[:SOURce]:LIST:DWELl:TYPE <val>{, <val>}
[:SOURce]:LIST:DWELl:TYPE LIST|STEP
[:SOURce]:LIST:DWELl:TYPE?
Edit Item
Press this softkey to change an item in the displayed list of sweep points. Use the
arrow keys or the front panel knob until the correct item is highlighted. Press
Edit
Item and the item is displayed in the active entry area. Change the value using the
up and down arrow keys, or the front panel knob and press the
Enter terminator
softkey. Or enter a new value using the numeric keypad and then press a units
terminator softkey.
A short cut for editing, Auto Edit, is also available. Highlight the item you want to
edit. Enter a new value using the numeric keys; editing immediately starts for the
highlighted item. Press the
Enter terminator softkey. The new value is stored and
the next item is highlighted.
Softkey Location: Press
Sweep/List, Configure List Sweep, Edit Item
Ext
This softkey is one of the choices in both the Sweep Trigger menu and the Point
Trigger menu. Press either the
Sweep Trigger softkey or the Point Trigger softkey
to view this key. Choosing
Ext allows you to trigger a sweep point-by-point on
either the negative or positive edge of a signal applied to the TRIGGER IN rear
panel connector. Set the polarity of the trigger signal using the
Trigger In Polarity
Neg Pos softkey.
Softkey Location: Press
Sweep/List, More (1 of 2), Sweep Trigger, Ext
or press Sweep/List, More (1 of 2), Point Trigger, Ext
Status after Normal Preset: Immediate