5-160
Hardkey and Softkey Reference
Sweep/List
Point Trigger
Pressing this softkey reveals a menu of choices for triggering a sweep
point-by-point. You can choose triggering that occurs after the dwell time for the
previous point has passed, triggering that is supplied by the HP-IB, triggering on
either the positive or negative edge of a signal supplied to the TRIGGER IN
connector, or triggering by the front panel
Trigger key.
Softkey Location: Press
Sweep/List, More (1 of 2), Point Trigger
Status after Normal Preset: Immediate
SCPI Commands:
[:SOURce]:LIST:TRIGger:SOURce BUS|IMMediate|EXTernal|KEY
[:SOURce]:LIST:TRIGger:SOURce?
Preset List
Press this softkey to eliminate the current sweep list and replace it with a new list
consisting of the following: one point at the maximum specified frequency, with
an amplitude of
135 dBm, and a dwell time of 2 ms. Be certain that you want to
delete the current sweep list; you cannot recover the information once you press
Preset List.
Softkey Location: Press
Sweep/List, Configure List Sweep, More (1 of 2),
Preset List
Single Sweep
Press this softkey to arm the sweep. The ARMED annunciator will turn on in the
display. The signal generator will begin the sweep as soon as the sweep trigger is
received. If you have set the sweep trigger to
Immediate, a sweep is initiated as
soon as you arm the sweep. The sweep sets the power and the frequency to the
first point of the sweep while it is not initiated or is waiting for the sweep trigger.
This softkey is disabled when
Sweep is set to Off.
Softkey Location: Press
Sweep/List, Single Sweep
SCPI Commands:
:INITiate:CONTinuous[:ALL] ON|OFF|1|0
:INITiate[:IMMediate][:ALL]
:INITiate[:IMMediate][:ALL]?