5-161
Hardkey and Softkey Reference
Sweep/List
Step Dwell
Press this softkey to set the dwell time for each point of a step sweep. To set the
dwell time, press
Step Dwell. The current dwell time is displayed in the active
entry area. Enter the desired dwell time using the up and down arrow keys, the
front panel knob, or enter a value using the numeric keypad and press the
sec or
msec terminator softkey. The range of values allowed is 1 ms through 60 s in 1 ms
increments. Dwell time is used when the point trigger is
Immediate. The dwell
time is the amount of time the sweep is guaranteed to pause after setting the
frequency and power for the current sweep/list point.
Softkey Location: Press
Sweep/List, Configure Step Sweep, Step Dwell
SCPI Commands:
[:SOURce]:SWEep:DWELl <val>
[:SOURce]:SWEep:DWELl?
Store to File
Press this softkey to store the current list sweep data to a file in internal
non-volatile memory. Press
Store To File and you will automatically enter the
editing mode to create a file name.
The file name can consist of up to 23 alpha-numeric and special characters. The
following steps show you how to create the following example file name,
FREQ
1453 TO 1465 MHZ:
NOTE: The following example uses softkeys to enter the file name. You can also use a combination
of the front panel knob and the arrow keys to accomplish the same function.
1. Press the softkey labeled
ABCDEFG. A menu is revealed containing a softkey for each
alphabetic character from A through G.
2. Press the softkey labeled
F. An F is entered as the first character in your file name.
Notice that the active entry area now shows Store to: F.
3. Enter an R in the file name by pressing the
OPQRSTU softkey and then pressing the R
softkey. Enter an E and a Q.
4. Next enter a space by pressing the
& # + softkey and then pressing .
5. Enter the value 1453 using the numeric keypad.
6. Enter another space and the letters TO.
7. Enter another space and the value 1465.
8. Enter another space and the letters MHZ.

Table of Contents