5-165
Hardkey and Softkey Reference
Sweep/List
SCPI Commands:
[:SOURce]:LIST:TRIGger:SOURce KEY
[:SOURce]:LIST:TRIGger:SOURce?
Trigger Out Polarity Neg Pos
This softkey toggles the polarity of the TTL signal that is output at the rear panel
TRIGGER OUT connector. When
Trigger Out Polarity Neg Pos is set to Pos the
output signal is asserted high (+5 V) at the start of a dwell sequence, or at the start
of waiting for the point trigger, and low (0 V) when the dwell is over, or when the
point trigger is received. Setting
Trigger Out Polarity Neg Pos to Neg reverses the
polarity where a high is 0 V and a low is +5 V.
This softkey is coupled to the
Trigger Out Polarity Neg Pos softkey in the
swept-sine AM, FM,
ΦM, and LF Out menus. When the polarity is toggled for
this softkey in any one of these menus, it affects all locations of the softkey.
Softkey Location: Press
Sweep/List, More (1 of 2), Trigger Out Polarity Neg Pos
Status after Normal Preset: Pos
SCPI Commands:
:TRIGger:OUTPut:POLarity POSitive|NEGative
:TRIGger:OUTPut:POLarity?