2-10
Using Functions
Creating a Step Sweep and a List Sweep
Creating a Step Sweep and a List Sweep
Using this procedure, you will learn two ways to set up the signal generator to
sweep a defined set of points. You will create a step sweep and then you will use
these points as the basis for a new list sweep.
In the first procedure, you will create a step sweep with the following ten equally
spaced points:
• frequency range from 525 MHz to 600 MHz
• power level from −20 dBm to 0 dBm
• dwell time 500 ms at each point
In the second procedure, you will take the step sweep points and edit several
points to change the sweep information.
Configuring a Step Sweep
1. Press Preset to return the signal generator to the factory-defined instrument state.
2. Press the front panel
Sweep/List key. The first level of sweep softkeys is displayed.
Toggle the
Sweep Type List Step softkey to Step.
3. Press the
Configure Step Sweep softkey. Another menu is displayed with softkeys
that you will use to create the sweep points.
4. Change the start frequency of the step sweep. Press the
Freq Start softkey. Enter
525 MHz using the numeric keypad and pressing the
MHz terminator softkey.
5. Change the stop frequency of the step sweep. Press the
Freq Stop softkey. Enter
600 MHz using the numeric keypad and pressing the
MHz terminator softkey.
6. Set the power level for the start of the step sweep. Press the
Ampl Start softkey. Enter
-20 dBm using the numeric keypad and pressing the
dBm terminator softkey.
7. Set the power level for the end of the step sweep. Press the
Ampl Stop softkey. Enter
0 dBm using the numeric keypad and pressing the
dBm terminator softkey.
8. Set the number of sweep points. Press the
#Points softkey. Enter 10 by rotating the
front panel knob until the number 10 is displayed.
9. Set the dwell time at each point. Press the
Step Dwell softkey. Enter 500 ms using the
numeric keypad and pressing the
msec terminator softkey.