2-5
Using Functions
Setting Frequency and Power Levels
6. Enter a 10 dB offset. Press the Ampl Offset softkey and enter 10 dB using the numeric
keypad. The amplitude display shows 20 dB (the power output by the hardware,
10 dBm, minus the reference power, 20 dBm, plus the offset, 10 dB). Notice that the
OFFS indicator is turned on. The power at the RF OUTPUT connector is still 10 dBm.