3-18
Troubleshooting
Execution Error
-202 Settings lost due to rtl
Indicates that a setting associated with a hard local control (see
IEEE 488.2, 5.6.15) was lost when the device changed to
LOCS from REMS or to LWLS from RWLS.
-210
Trigger error
Indicates that a GET, *TRG, or a triggering signal could not
be executed due to an error.
-211
Trigger ignored
Indicates that a GET, *TRG, or triggering signal was received
and recognized by the device, but was ignored because of
device timing considerations. For example, the device was not
ready to respond.
-212
Arm ignored
Indicates that an arming signal was received and recognized by
the device but was ignored.
-213
Init ignored
Indicates that a request for a measurement initiation was
ignored as another measurement was already in progress.
-214
Trigger deadlock
Indicates that a trigger source for the initiation of a
measurement is set to
GET and a subsequent measurement
query is received. The measurement cannot begin until a
GET is
received, but the
GET would cause an INTERRUPTED error.
-215
Arm deadlock
Indicates that the arm source for the initiation of a measurement
is set to
GET and a subsequent measurement query is received.
The measurement cannot begin until a
GET is received, but the
GET would cause an INTERRUPTED error.
-220
Parameter error
Indicates that a program data element related error has
occurred. This particular error message is used if the device
cannot detect a more specific errors described for errors -221
through -229.