3-49
Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
State Save Recall Error;Recall non-saved state
register. Recall ignored.
Indicates that a recall was attempted for a state register that is
unused. If state file exists, delete explicitly using catalog.
State Save Recall Error;Delete sequence
<sequence_name> ignored.
Indicates that a STATE: file in a sequence that is being deleted
could not be deleted due to the file not being found, data
corruption, etc. If state file exists, delete explicitly using the
memory catalog.
State Save Recall Error;The state file is from
a different firmware revision that does not
support comments.
Indicates that an attempt was made to write a comment to a
state file revision that does not support comments. Comments
in saved state files are not supported by the A.01.00 and
A.01.01 releases of the instrument firmware.