6 - 8
HAPSITE ER Operating Manual
5 When the AutoTune is finished, the following message will be displayed. Click
OK. (See Figure 6-14.)
Figure 6-14 AutoTune Complete
NOTE: If AutoTune fails, refer to section 9.4, Performing Manual Tune, on
page 9-7.
6 A concentrator cleanout will also be run as part of the preparation of HAPSITE.
Hold the probe in a clean environment for the duration of the cleanout. If the
cleanout is successful, a SYSTEM IS READY message will be displayed on the
front panel. The TIC on the chromatogram will be less than 5 million.
NOTE: If the cleanout is unsuccessful, refer to section 4.1.3, Concentrator
Cleanout Failure, on page 4-11.
NOTE: If this method is not the desired method, refer to section 4.2, Selecting
a Different Method Using the SELECT METHOD Icon, on page 4-13.
7 Once all temperature zones have reached their setpoints, a prompt will be
displayed to Press RUN to start method.
8 Click RUN button on the pop-up window or from the Control Panel on the
screen. (See Figure 6-15.)
Figure 6-15 Run Button
9 When the HAPSITE screen prompts Collect Sample Now, place the probe
over sample for the entire specified sampling time. Be sure to keep the probe
over the sample for both the Line Purge and Concfill events. (See Figure 6-16.)
NOTE: The Line Purge event is collected by time and the Concfill is collected
by volume.
CAUTION
Do not place the sample probe in liquids while sampling.