6 - 10
HAPSITE ER Operating Manual
12 The message METHOD FINISHED will appear on the HAPSITE screen when
the run is complete. (See Figure 6-18.) An example of a completed
chromatogram on the laptop is shown in Figure 6-19.
Figure 6-18 Method Finished for Concentrator
Figure 6-19 Concentrator Run Finished
13 Review results at the end of the run. If red lines appear on the chromatogram,
saturation has occurred. To clear saturation, run blank runs until the saturation
has cleared.
CAUTION
The concentrator feature has increased sensitivity. Take
care to avoid saturating HAPSITE.