8 - 32
HAPSITE ER Operating Manual
On the Show/Update Current Peaks window, the highlighted peak will be
displayed with the dots along the peak representing each individual scan. The
following will also be displayed:
Number of Peaks . . . . . . . . . . . . . . The number of peaks that have been
identified by the Show/Update Current
Peaks program.
Retention Time . . . . . . . . . . . . . . . . The time the compound elutes from the
column.
Scan Range. . . . . . . . . . . . . . . . . . . The range of scans that encompass the
peak.
Area . . . . . . . . . . . . . . . . . . . . . . . . . The area of the peak.
Percent Area . . . . . . . . . . . . . . . . . . The ratio of the TIC of the peak to the total
TIC multiplied by 100.
Add Peaks . . . . . . . . . . . . . . . . . . . . The peaks that were not automatically
located by the software can be added to the
Show/Update Current Peaks window using
the Range Tool. See section 8.7.2, Range
Tool, on page 8-42.
Hide Graph . . . . . . . . . . . . . . . . . . . Checking the Hide Graph box will remove
the graph from the Show/Update Current
Peaks window.
Redo Peak Search . . . . . . . . . . . . . The software will clear the identifications
from the Hits Found column.
Peak Search Parameters . . . . . . . . See section 10.10.2, Setting Up a
Quantitative Search, on page 10-43 for more
information.
NIST Search Setup . . . . . . . . . . . . . Allows the pathway for the NIST libraries to
be set. The library pathways are set at the
factory, but may need reset if the software is
reloaded. See section 8.7.0.3, NIST Search
Setup, on page 8-33 for instructions.
Report Preview . . . . . . . . . . . . . . . . The information in the Show/Update
Current Peaks is displayed in a text format.
See section 8.7, Show/Update Current
Peaks, on page 8-28 for instructions.