10 - 42
HAPSITE ER Operating Manual
Advanced Settings
Figure 10-47 Advanced Settings
NOTE: INFICON does not recommend changing the Advanced Settings.
Noise Thresholding Level . . . . . . . Refers to the minimum signal recorded. Will
filter out noise along the baseline.
Chromatogram Peak Width . . . . . . Deconvoluted peaks will maintain same
shape, because the width, in amus, has been
specified.
Reject Bin . . . . . . . . . . . . . . . . . . . . Rejects peaks that have less than a set
number of scans.
Deconvolution Level . . . . . . . . . . . As the level of deconvolution increases, the
software increases the separation between
peaks. The default setting is medium, but low
and high options are available.
Peak Shape . . . . . . . . . . . . . . . . . . . The shape requirement allows all of the
deconvoluted peaks to maintain the same
shape. As the shape requirement increases,
the shape of the individual ions will be more
uniform.
The Search Trigger Criteria section of the Search page determines when an
AMDIS search will be run. There are three choices, see Figure 10-48.
Figure 10-48 Search Trigger Criteria
Every ____ Scans . . . . . . . . . . . . . . Determines the scan interval for running an
AMDIS search. The default value for this is
100 scans.