3 - 23
HAPSITE ER Operating Manual
5 Lift the top elbow. Gently lift and angle the concentrator out of the fixture. (See
Figure 3-42.)
Figure 3-42 Inserting the Concentrator
The elbow fittings may be hot. Allow for these fittings to
cool before continuing.
6 Carefully, remove the concentrator from the bottom elbow. Avoid losing the
ferrules which are located inside the nuts. (See Figure 3-43.)
Figure 3-43 Removing the Concentrator
7 Dispose of the damaged concentrator according to local procedure. Please be
aware that the concentrator contains glass.
The concentrator contains glass. Wear appropriate
personal protection equipment when handling a broken
concentrator.
8 Remove the new concentrator from the storage vial and unwrap it.
9 Ensure that a Teflon
®
ferrule is installed into the threaded end of each plastic
nut with the wide end of the cone facing toward the center of the concentrator.
(See Figure 3-44.)