3 - 25
HAPSITE ER Operating Manual
11 The Tenax Concentrator does not have a specific orientation.
(See Figure 3-45.)
Figure 3-45 Tenax Concentrator
12 Place the Tri-Bed concentrator:
12a Tri-Bed concentrator: While holding the nut and ferrule in place, carefully
place the smooth metal sleeve end of the Tri-Bed concentrator into the
lower elbow fitting.
12b Tenax concentrator: While holding the nut and ferrule in place, carefully
place either end of the concentrator into the lower elbow fitting. (See
Figure 3-46.)
Figure 3-46 Placing Concentrator in the Bottom Elbow