4 - 18
HAPSITE ER Operating Manual
Procedure
1 If an Analyze method is going to be run after Survey, verify that the appropriate
sample configuration (i.e., concentrator) is installed.
2 When powered on or taken out of Extended Standby, HAPSITE ER will
automatically start preparing an ER Survey and Analyze method if the probe is
attached. Refer to section 4.1, Starting HAPSITE ER in Portable Mode, on page
4-1.
3 Verify that the desired Analyze method is listed under the Survey method. In
Figure 4-30, the Analyze method is ER_Air_Tri-Bed_PPB_Standard.
Figure 4-30 Analyze Method
4 A SYSTEM READY message will be displayed with a prompt to press Survey
or Analyze when HAPSITE ER is ready to sample. (See Figure 4-31.)
Figure 4-31 System Ready