4 - 20
HAPSITE ER Operating Manual
8 Sample air away from the point of concern for one minute. Remember to note
the background TIC. (See Figure 4-35.)
Figure 4-35 Background Sampling
9 Hold the probe over the sample of interest for up to 1 minute. A peak may
appear if the compound present is greater than 1 ppm. A compound
identification may also be present on the HAPSITE ER screen. (See
Figure 4-36.)
Figure 4-36 Survey
CAUTION
Do not touch the sample with the probe.
Do not allow liquids to enter the probe.