2- 76 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Command Summary
Using Advanced Functions
:SYST:PERS[:MAN] maker
:SYST:PERS[:MAN]? [maker]
Changes the manufacturer name of the
*IDN? command response.
maker=DEF|AGIL|KEYS or a quoted
string
Query does not support a quoted string.
:SYST:PERS[:MAN]:DEF
Clears the :SYST:PERS[:MAN]
command setting and applies the default
value “Keysight Technologies.”
:SYST:PERS:MOD model
:SYST:PERS:MOD?
Changes the model number of the *IDN?
command response.
model: a quoted string
:SYST:PERS:MOD:DEF
Clears the :SYST:PERS:MOD command
setting and applies the default value (the
model number on the front panel).
:SYST:PON memory
Specifies the power-on state.
memory=RST|RCL0|RCL1|RCL2|RCL3|
RCL4
RST
:SYST:PRES
Presets the instrument settings and the
front panel display.
:SYST:SET data
:SYST:SET?
Sends or loads the instrument setup data.
data: Instrument setup data. Parameter
data type is a definite length arbitrary
binary block.
:SYST:TIME hour,minute,second
:SYST:TIME?
Sets the time of the internal clock.
hour: Hour. Integer from 0 to 23.
minute: Minute. Integer from 0 to 59.
second: Second. Integer from 0 to 59.
:SYST:TIME:TIM:COUN?
Returns the present count of the timer.
Command Summary Reset setting

Table of Contents