Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 113
Subsystem Commands
SOURce Subsystem
[:SOURce]:ARB:<CURRent|VOLTage>:SQUare:TOP[:
LEVel]
Sets the top level of the current or voltage square waveform output.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SQUare:TOP[:LEVel] data
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SQUare:TOP[:LEVel]? [data]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter data Square waveform top level. value (see “Source Output Ranges
on page 2-36) |MINimum|MAXimum|DEFault (default is 0.0).
Parameter data type is NRf+. Query does not support
data=value.
Query response response <newline>
response returns the present setting. If a parameter is specified, response returns the
value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :ARB:CURR:SQU:TOP 1E-3
:SOUR2:ARB:CURR:SQU:TOP?
[:SOURce]:ARB:<CURRent|VOLTage>:SQUare:TOP:
TIME
Sets the top time of the current or voltage square waveform output. Waveform
period must not exceed 1000 seconds.
Syntax [:SOURce[c]]:ARB:<CURRent|VOLTage>:SQUare:TOP:TIME time
[:SOURce[c]]:ARB:<CURRent|VOLTage>:SQUare:TOP:TIME? [time]
For <CURRent|VOLTage>, specify CURRent for current output or VOLTage for
voltage output.
Parameter time Square waveform top time, in seconds. value (0 to 1000)|
MINimum|MAXimum|DEFault (default is 0). Parameter data
type is NRf+. Query does not support time=value.
Query response response <newline>

Table of Contents