EasyManuals Logo

Keysight Technologies B2961A Command Reference

Keysight Technologies B2961A
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #221 background imageLoading...
Page #221 background image
Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 83
Subsystem Commands
READ Subsystem
This example shows the data containing the current data (ch1currN) of the 10-step
sweep measurement by channel 1, and the current data (ch2currN) of the 5-step
sweep measurement by channel 2.
If the measurement function is not enabled or no data exists, response returns
+9.910000E+37 (ASCII) or NaN (IEEE-754) which indicates “not a number”.
Example :READ:ARR:CURR? (@2,1)
:READ[:SCALar]?
Executes the :INITiate command and the :FETCh[:SCALar]? command in series,
and returns the latest voltage measurement data, current measurement data,
resistance measurement data, time data, status data, or source output setting data
specified by the
:FORMat:ELEMents:SENSe command. The data is not cleared
until the :INITiate, :MEASure, :READ, or :SENSe:DATA:CLEar command is
executed.
Syntax :READ[:SCALar]? [chanlist]
Parameter chanlist Channels to return the data. Parameter data type is channel list.
(@1)|(@2)|(@1,2)|(@1:2)|(@2,1)|(@2:1). See
“Channel List
Parameter” on page 1-8.
(@1) selects channel 1 only. (@2) selects channel 2 only. (@1,2), (@1:2), (@2,1),
and (@2:1) selects both channels 1 and 2.
If this parameter is not specified, chanlist=(@1) is set.
Query response response <newline>
response returns the latest data specified by the :FORMat:ELEMents:SENSe
command. Response data type is NR3. See “Data Output Format” on page 1-12.
If both channels 1 and 2 are selected by chanlist, response returns the channel 1 data
and the channel 2 data in this order. See the following example. With the ASCII data
output format, each data is separated by a comma.
ch1curr10,ch1sour10,ch2curr5,ch2sour5
This example shows the data containing the latest current data (ch1curr10) and
source data (ch1sour10) of the 10-step sweep measurement by channel 1, and the
latest current data (ch2curr5) and source data (ch2sour5) of the 5-step sweep
measurement by channel 2.

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals