5- 12 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Error Messages
B2961A/B2962A Specific Error
B2961A/B2962A Specific Error
Positive error numbers are Keysight B2961A/B2962A specific errors, not standard
SCPI errors. Consult service for errors 111 to 140 and 142.
Error 101 Wrong password
Error 102 Enter password for calibration
Error 103 Data load failed
Error 104 Data save failed
Error 111 Self-calibration failed; Voltage offset, item; channel n
Failed the voltage offset self-calibration specified by item and channel n.
Error 112 Self-calibration failed; Current offset, item; channel n
Failed the current offset self-calibration specified by item and channel n.
Error 113 Self-calibration failed; Voltage gain, item; channel n
Failed the voltage gain self-calibration specified by item and channel n.
Error 114 Self-calibration failed; Current gain, item; channel n
Failed the current gain self-calibration specified by item and channel n.
Error 115 Self-calibration failed; CMR DAC, item; channel n
Failed the CMR DAC self-calibration specified by item and channel n.
Error 121 Self-test failed; CPU communication, item; channel n
Failed the CPU communication test specified by item and channel n.
Error 122 Self-test failed; Fan status, item; channel n
Failed the fan status test specified by item and channel n.
Error 131 Self-test failed; Module communication, item; channel n
Failed the module communication test specified by item and channel n.