Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 99
Subsystem Commands
SENSe Subsystem
Query response gain <newline>
gain returns the present setting of the gain value. If a parameter is specified, gain
returns the value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :SENS:WAIT:GAIN 0.5
:SENS2:WAIT:GAIN?
:SENSe:WAIT:OFFSet
Sets the offset value used for calculating the measurement wait time for the
specified channel. See
:SENSe:WAIT[:STATe].
Syntax :SENSe[c]:WAIT:OFFSet offset
:SENSe[c]:WAIT:OFFSet? [offset]
Parameter offset value (0 to 1 seconds)|MINimum|MAXimum|DEFault (default
is 0). Parameter data type is NRf. Query does not support
offset=value.
Query response offset <newline>
offset returns the present setting of the offset value. If a parameter is specified, offset
returns the value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :SENS:WAIT:OFFS 0.5
:SENS2:WAIT:OFFS?
:SENSe:WAIT[:STATe]
Enables or disables the measurement wait time for the specified channel. The wait
time is defined as the time the measurement channel cannot start measurement after
the start of a DC output or the trailing edge of a pulse.
Syntax :SENSe[c]:WAIT[:STATe] mode
:SENSe[c]:WAIT[:STATe]?
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables the measurement wait time. The wait time is set to 0.
mode=1 or ON enables the measurement wait time given by the following formula.