4- 200 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Commands
TRIGger Subsystem
source=INT1 or INT2 selects a signal from the internal bus 1 or 2, respectively.
source=LANn selects a LXI trigger specified by the
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:SOURce:LAN command.
source=EXTn selects a signal from the GPIO pin n, which is an input port of the
Digital I/O D-sub connector on the rear panel. n=1 to 14.
Query response response <newline>
response returns the present setting of trigger source, AINT, BUS, TIM, INT1,
INT2, LAN, or EXT1 through EXT14. Response data type is CRD.
Example :TRIG:SOUR EXT1
:TRIG2:TRAN:SOUR:SIGN?
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:TIMer
Sets the interval of the TIMer trigger source for the specified device action.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:TIMer interval
:TRIGger[c]<:ACQuire|:TRANsient>:TIMer? [interval]
:TRIGger[c][:ALL]:TIMer? interval
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.
Parameter interval Interval, in seconds. value (1E-5 to 1E+5)|MINimum|
MAXimum|DEFault (default is 1E-5). Parameter data type is
NRf+. Query does not support interval=value.
Query response response <newline>
response returns the present setting of interval. If a parameter is specified, response
returns the value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :TRIG:TIM 2E-4
:TRIG2:TRAN:TIM?