Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 179
Subsystem Commands
SYSTem Subsystem
name: a quoted string (alpha numeric and whitespace) for
manufacturer name
Query response maker <newline>
If no parameter is specified, maker returns a quoted string of the present setting. If a
parameter is specified, maker returns a quoted string associated with the parameter
shown above. Response data type is SRD.
Example :SYST:PERS AGIL
:SYST:PERS?
:SYSTem:PERSona[:MANufacturer]:DEFault
Clears the :SYSTem:PERSona[:MANufacturer] command setting and applies the
default value “Keysight Technologies.”
The setting change is effective after rebooting the B2961A/B2962A.
Syntax :SYSTem:PERSona[:MANufacturer]:DEFault
Example :SYST:PERS:DEF
:SYSTem:PERSona:MODel
Changes the model number of the *IDN? command response.
The setting change is effective after rebooting the B2961A/B2962A. The model
number is set back to the default value (the model number on the front panel) by the
:SYSTem:PERSona:MODel:DEFault command or by pressing the front panel
System > More > Factory function keys which applies the factory shipment initial
condition. This command setting is not changed by power off or the *RST
command.
NOTE If you change the model number, instrument control software (e.g. Quick IV,
IVI-COM/C, LabVIEW, and so on) may not detect this instrument.
Syntax :SYSTem:PERSona:MODel model
:SYSTem:PERSona:MODel?
Parameter model A quoted string (alpha numeric) for model number. Parameter
data type is SPD.