Keysight B2961A/B2962A SCPI Command Reference, Edition 6 4- 67
Subsystem Commands
OUTPut Subsystem
NOTE This command setting is not applied to the output-off process triggered by the
emergency condition such as the over voltage/current protection, interlock open, and
over temperature protection. Then the output voltage is immediately set to 0 V and
the output switch is set to off.
Query response mode <newline>
mode is NORM, HIZ, or ZERO, and indicates the source condition after output off.
Response data type is CRD.
Example :OUTP:OFF:MODE HIZ
:OUTP2:OFF:MODE?
:OUTPut:ON:AUTO
Enables or disables the automatic output on function.
Syntax :OUTPut[c]:ON:AUTO mode
:OUTPut[c]:ON:AUTO?
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables the automatic output on function.
mode=1 or ON enables the automatic output on function. If this function is enabled,
the source output is automatically turned on when the :INITiate or :READ command
is sent.
Query response mode <newline>
mode is 0 or 1, and indicates that the automatic output on function is off or on,
respectively. Response data type is NR1.
Example :OUTP:ON:AUTO 0
:OUTP2:ON:AUTO?
:OUTPut:PROTection[:STATe]
Enables or disables the over voltage/current protection. If this function is enabled,
the source channel sets the output to 0 V and sets the output switch to off
automatically and immediately when it reaches the compliance status.